CMG-DM24S12AMS Operator's guide

11.Calibrate ports CH3 and CH4 in the same way, but with the N/S

calibration channel selected. You must wait for the previous

calibration to finish before doing this: clicking Inject now has no
effect whilst the Calibration enable relay is open.

12.Calibrate ports CH5 and CH6 in the same way, but with the E/W

calibration channel selected.

13.Close the window and right-click on the second internal

digitiser's icon. Select Control... and open the Calibration pane.

14.Repeat steps 2 – 11 to calibrate ports CH7 – 12 in pairs: CH7 and

CH8 on the Z channel, CH9 and CH10 on the N/S channel, and
CH11 and CH12 on the E/W channel.

If you prefer, you can inject your own signals into the system at any
point (together with a Calibration enable signal, if required) to provide
independent measurements, and to check that the voltages around the

calibration loop are consistent. For reference, a DM24-series digitiser

will generate a calibration signal of around 16000 counts / 4 V when set

to 100% (sine-wave or step), and around 10000 counts / 2.5 V when set to

50%.

70

Issue B