The STA and LTA columns allow you to set the intervals over which the
two averages are calculated, in seconds. Typically, the time interval for

the short term average should be about as long as the signals you want

to trigger on, while the long term average should be taken over a much
longer interval. Both the STA and LTA values are recalculated

continually, even during a trigger.

The Ratio column determines by what factor the STA and LTA must
differ for the trigger to be passed. Finding the ratio most suited to your
needs is best done by experiment. Too high a value will result in events

being missed, while too low a value will result in spurious non-seismic

noise triggering the system. Like the averages, their ratio is continuously

recalculated for all components. Note that none of the boxes are allowed

to be empty, and so you will need to enter the new value before

removing the old one. Alternatively, you can use the up and down
cursor keys to change the values.

For example, setting the STA to 1 second, the LTA to 10 seconds and the
Ratio to 4 would give rise to the following trigger behaviour: