Guralp Systems CMG-DM24S12AMS User Manual
Page 30

CMG-DM24S12AMS Operator's guide
a hexadecimal integer (i.e. 0 – 9 and A – F for 10 through 15). If a Güralp
digital instrument is attached to a DIGITAL A – F input, the Z, N/S and
E/W Mass Position Mux channels appear as M8, M9 and MA respectively.
The unit's built-in DM24 digitisers do not report mass positions,
although you can access the user-defined channels 0 – 7 for both
through the AUXILIARY connector on the casing. Digitisers also
monitor their internal temperature, which is reported on channel ME.
Ports
Scream! can normally auto-detect the settings it needs to communicate
with any instruments you connect to the DM24S12AMS. However, if
you connect a digitiser to the unit through a telemetry link or a Güralp
Storage and Acquisition Module (SAM), you may need to set the baud
rate and other COM port parameters to match those of the link or
module you are using.
The Baud Rates pane of the Configuration setup window allows you to
program the baud rate and stop bits for the digitiser's COM port.
30
Issue B