AMETEK SLM-1 User Manual
Page 37

Sorensen SLM-1 Chassis
GPIB/RS-232 Programming
Command Syntax:
All Modules:
[GLOB:] LEVE {SP}{HIGH|LOW}{NL}
Description:
LEVE LOW is Set LOW current level in CC mode, LOW resistance level in CR mode, or
LOW voltage level in CV mode at the active load channel.
LEVE 1 is Set HIGH current level in CC mode, HIGH resistance level in CR mode, or HIGH
voltage level in CV mode at the active load channel.
PRESET ON/OFF
Purpose:
Set the upper or lower digit multi-function meter to display the programming load level.
Command Syntax:
All Modules:
[GLOB:]PRES{SP}{0|1|OFF|ON}{NL}
Description:
GLOB:PRES ON is set all the load module in the SLM chassis to preset on status.
MODE
Purpose:
Select the operating mode of Electronic load module.
Command Syntax:
All Modules:
[GLOB:]MODE{SP}{0|1|2|3|CC|CR|CV|CP}{NL}
Description:
GLOB:MODE CC ; set the presently operating mode to Constant Current mode for all load
module in the SLM chassis.
MODE CV ; set the presently operating mode to Constant Voltage mode.
MODE 1 ; set the presently operating mode to Constant Resistance mode.
MODE CP ; set the presently operation mode to Constant Power mode.
Note
:
MODE CV is available in DC loads only.
MODE CP is available in DC, single input loads only.
CLEAR status register
Purpose:
CLEar the PROT and ERR status byte registers.
Command Syntax:
All Modules:
CLER{NL}
Description:
CLER ; clear the PROT and ERR status byte register, the PROT and ERR status byte
register will indicate “0” after executing the CLER command.
STORE
Purpose:
STORE the load level and load status into the memory of the SL-series electronic LOAD.
Command Syntax:
SLM DC, SLD Modules: STOR{SP}{m[,n]}{;|NL}
SLM AC Modules:
STOR{SP}{m}{;|ML}
Description:
M540070-01 Rev B
4-11