beautypg.com

Connections – AMETEK SLM-1 User Manual

Page 19

background image

Sorensen SLM-4 Chassis

Configuration

2.1.3 Environmental Requirements

For indoor use only

Installation Category II (over voltage)

Pollution Degree 2

Altitude up to 2000 meters (with power derating)

Relative Humidity 80% RH Max

Ambient Operating Temperature 0-40°C with ideal between 25°C ± 5°C

2.1.4

Service or Repair

If the instrument is damaged, please attach a tag to the instrument, identifying the
owner and indicating the required service. Follow the procedures detailed in the
warranty at the end of this manual.

2.1.5 Accessories

The following parts should be included in the shipment:

• Three-conductor 18 gauge power cord, 1 ea.
• SLM-1 Electronic Load Chassis operation manual, 1 ea. (this manual)

If a module is configured with the chassis, the following parts should be included in
the shipment:

• Vsense cable(s) for each input
• Hook terminals for each input
• Banana plugs for each input
• Imonitor BNC cable (if applicable)

2.2

CONNECTIONS

2.2.1 GPIB

Connection

The GPIB connector on the rear panel (Figure 2-3) connects the SLM-1 chassis to a
controller and to other GPIB devices. A GPIB system can be connected in any
configuration (star, linear, or both), with the following provisions:

• The maximum number of devices including the controller is no more than 15.
• The maximum length of all cables is no more than 2 meters times the number

of devices connected together, up to 20 meters maximum.

Ensure that the lock screws are firmly hand-tightened; use a screwdriver only for the
removal of screws. Figure 2-3 shows the rear panel of the SLM-1 chassis, where the
GPIB connector is located. The GPIB address of the SLM-1 chassis is set on the
front panel (Figure 3-1).

M540070-01 Rev B

2-3