Sorensen SLM-1 Chassis

GPIB/RS-232 Programming

REMARKS:

1. d : 0 - 9
2. GLOB : GLOBAL (ALL CHANNELS ACTIVE AT SAME TIME)
3. CURRENT ENGINEERING UNIT : A
4. VOLTAGE ENGINEERING UNIT : V
5. RESISTANCE ENGINEERING UNIT :

Ω

6. PERIOD ENGINEERING UNIT : mS
7. SLEW-RATE ENGINEERING UNIT : A/

μS


Note

: The RS-232 command set is the same as the GPIB command set.

4.4

GPIB/RS-232 COMMAND DESCRIPTION

4.4.1 Setting

Commands

CHANNEL

Purpose:
”CHAN” selects the multiple Electronic load channel to which all subsequent channel-
specific commands will be directed.
Command Syntax:
All Modules:

CHAN{SP}n{;NL}

Description:
”CHAN” command selects the specified Electronic load Channel A or B as the Electronic
load module number (from left to right). This command is a channel independent command;
therefore, this command should be programmed before an electronic load channel
dependent

command.

The load channel numbercommand applies to SLD modules only..
Example:

CHAN A selects channel A of an installed SLD module.
Note:

Please refer to Appendices C, D and E for proper programming procedure of SL-

series electronic load modules.

CURRENT

Level

Purpose:
The load current setting in Constant Current mode.
Command Syntax:
All SLM Modules:

CC:{LOW|HIGH}{SP}{NR2}{;|NL}

SLD Modules:

CC{SP}{NR2}{;|NL}

Description:
CC:{LOW|HIGH}{SP}{NR2}{;|NL}
Sets the current level of SLM-series AC or DC Electronic Load modules.

CC:{SP}{NR2}{;|NL}
This command is used to set the load current level for CC static mode of SLD-series
electronic load module.

M540070-01 Rev B