Gpib/rs-232 command description – AMETEK SLM-1 User Manual
Page 33
Sorensen SLM-1 Chassis
GPIB/RS-232 Programming
REMARKS:
1. d : 0 - 9
2. GLOB : GLOBAL (ALL CHANNELS ACTIVE AT SAME TIME)
3. CURRENT ENGINEERING UNIT : A
4. VOLTAGE ENGINEERING UNIT : V
5. RESISTANCE ENGINEERING UNIT :
Ω
6. PERIOD ENGINEERING UNIT : mS
7. SLEW-RATE ENGINEERING UNIT : A/
μS
Note
: The RS-232 command set is the same as the GPIB command set.
4.4
GPIB/RS-232 COMMAND DESCRIPTION
4.4.1 Setting
Commands
CHANNEL
Purpose:
”CHAN” selects the multiple Electronic load channel to which all subsequent channel-
specific commands will be directed.
Command Syntax:
All Modules:
CHAN{SP}n{;NL}
Description:
”CHAN” command selects the specified Electronic load Channel A or B as the Electronic
load module number (from left to right). This command is a channel independent command;
therefore, this command should be programmed before an electronic load channel
dependent
command.
The load channel numbercommand applies to SLD modules only..
Example:
CHAN A selects channel A of an installed SLD module.
Note:
Please refer to Appendices C, D and E for proper programming procedure of SL-
series electronic load modules.
CURRENT
Level
Purpose:
The load current setting in Constant Current mode.
Command Syntax:
All SLM Modules:
CC:{LOW|HIGH}{SP}{NR2}{;|NL}
SLD Modules:
CC{SP}{NR2}{;|NL}
Description:
CC:{LOW|HIGH}{SP}{NR2}{;|NL}
Sets the current level of SLM-series AC or DC Electronic Load modules.
CC:{SP}{NR2}{;|NL}
This command is used to set the load current level for CC static mode of SLD-series
electronic load module.
M540070-01 Rev B
4-7