SECTION 2

CALIBRATION PROCEDURE

2.1 CALIBRATION

SETUP

Prior to starting the calibration procedure, gather the necessary equipment
(Section 2.1.1), then power up the chassis and allow it to warm up for 15
minutes.

2.1.1

E

QUIPMENT

R

EQUIRED

Voltage/Current Source/Calibrator: Krohn-Hite EDC 521.

Current Shunt: Prodigit 7550.

Digital Multi-meter (DMM): Agilent 34401A.

DC Power Supply: SGA80-125. (Supply may depend upon models

being calibrated. This model covers any module in the SLD
family.)

Current Probe: Tektronix A6303.

2.2 CHANNEL

A

CALIBRATION

Press CHAN key to light the A LED.

Press both MODE and LIMIT keys simultaneously to enter calibration mode.
Initial default mode is Constant Current (CC) Mode.

SLD Calibration Manual

2-1