Operation Description

Elgar Electronics Corporation

1.1.1

N

AVIGATION AND

S

ETTING

1.

Press CHAN key to select channel (A or B) to be calibrated (each
channel is calibrated separately).

2.

Press MODE and LIMIT keys simultaneously to enter calibration
mode (Constant Current is default initial mode).

3.

Press

ƒ key to return to previous item (parameter); press

„

to

advance to next item
NOTE: There are 18 items (parameters) to calibrate.

4.

Load level and status are set to default value automatically at each
step.

5.

Press

…† keys to adjust the calibration values.

6.

Press DUAL V/A key to store the calibration data.

7.

Press LIMIT key to exit calibration mode.

Refer to Figure 1-2 for a flowchart of the calibration procedure.

Figure 1-2. SLD Series Calibration Procedure Flowchart

1-2

SLD Calibration Manual