Calibration Procedure

Elgar Electronics Corporation

CR Range II Offset and Gain, Channel A CR Mode

1.

Set load to match the value in CR Range I Offset column of
Table 2-21, for the model being calibrated.

2.

Set the DC voltage standard to match the value in IV Sense Input
column of Table 2-21 for the same model being calibrated.

3.

LOAD ON (press LOAD key until its LED is lit).

4.

Adjust using

…† keys until DMM reading matches the value in DMM

Reading column of Table 2-21.

ITEM

MODEL

CR Range II OFFSET V SENSE input DMM Reading

SLD-60-505-255

12.00

60.00 V

5.00 A

SLD-61-505-255

12.00

60.00 V

5.00 A

SLD-60-20-102

3.000

15.00 V

5.00 A

SLD-80-20-102

4.000

15.00V 3.750

A

SLD-61-5-752

12.000

60.00 V

5.000 A

SLD-62-5-752

12.000

60.00 V

5.000 A

17

SLD-60-105-550

12.000

60.00 V

5.000 A

Table 2-21. Channel B CR Mode Range II Offset Calibration Settings

5.

Scroll to next calibration parameter, Item 18 (press

„).

6.

Set the DC voltage standard to match the value in V Sense Input
column of Table 2-22 for the same model being calibrated.

7.

LOAD ON (press LOAD key until its LED is lit). Adjust using

…† keys

until DMM reading matches the value in DMM Reading column of
Table 2-22.

8.

Repeat Steps 1 through 7 (press

ƒ to return to previous item) until no

further adjustments are necessary for the DMM reading to match the
value in the DMM Reading column for both calibration parameters
(Items 17 and 18).

ITEM

MODEL

CR Range II GAIN

V SENSE input DMM Reading

SLD-60-505-255

2.000

5.000 V

2.500 A

SLD-61-505-255

2.000

5.000 V

2.500 A

SLD-60-20-102

0.500

5.000 V

10.00 A

SLD-80-20-102

0.133

2.66 V

20.00 A

SLD-61-5-752

2.000

5.000 V

2.500 A

SLD-62-5-752

2.000

5.000 V

2.500 A

18

SLD-60-105-550

2.000

5.000 V

2.500 A

Table 2-22. Channel B CR Mode Range II Gain Calibration Settings

2-24

SLD Calibration Manual