AMETEK SL Series Software User Manual
Page 32

Calibration Procedure
Elgar Electronics Corporation
CR Range II Offset and Gain, Channel A CR Mode
1.
Set load to match the value in CR Range I Offset column of
Table 2-21, for the model being calibrated.
2.
Set the DC voltage standard to match the value in IV Sense Input
column of Table 2-21 for the same model being calibrated.
3.
LOAD ON (press LOAD key until its LED is lit).
4.
Adjust using
keys until DMM reading matches the value in DMM
Reading column of Table 2-21.
ITEM
MODEL
CR Range II OFFSET V SENSE input DMM Reading
SLD-60-505-255
12.00
Ω
60.00 V
5.00 A
SLD-61-505-255
12.00
Ω
60.00 V
5.00 A
SLD-60-20-102
3.000
Ω
15.00 V
5.00 A
SLD-80-20-102
4.000
Ω
15.00V 3.750
A
SLD-61-5-752
12.000
Ω
60.00 V
5.000 A
SLD-62-5-752
12.000
Ω
60.00 V
5.000 A
17
SLD-60-105-550
12.000
Ω
60.00 V
5.000 A
Table 2-21. Channel B CR Mode Range II Offset Calibration Settings
5.
Scroll to next calibration parameter, Item 18 (press
).
6.
Set the DC voltage standard to match the value in V Sense Input
column of Table 2-22 for the same model being calibrated.
7.
LOAD ON (press LOAD key until its LED is lit). Adjust using
keys
until DMM reading matches the value in DMM Reading column of
Table 2-22.
8.
Repeat Steps 1 through 7 (press
to return to previous item) until no
further adjustments are necessary for the DMM reading to match the
value in the DMM Reading column for both calibration parameters
(Items 17 and 18).
ITEM
MODEL
CR Range II GAIN
V SENSE input DMM Reading
SLD-60-505-255
2.000
Ω
5.000 V
2.500 A
SLD-61-505-255
2.000
Ω
5.000 V
2.500 A
SLD-60-20-102
0.500
Ω
5.000 V
10.00 A
SLD-80-20-102
0.133
Ω
2.66 V
20.00 A
SLD-61-5-752
2.000
Ω
5.000 V
2.500 A
SLD-62-5-752
2.000
Ω
5.000 V
2.500 A
18
SLD-60-105-550
2.000
Ω
5.000 V
2.500 A
Table 2-22. Channel B CR Mode Range II Gain Calibration Settings
2-24
SLD Calibration Manual