Calibration Procedure

Range I Offset and Gain, Channel B CC Mode

1.

Set DC power supply to +5VDC.

2.

PRES ON (press PRES key until its LED is lit); set load current setting
to match the value shown in Load Setting column of
Table 2-12
, for the model being calibrated.

3.

LOAD ON (press LOAD key until its LED is lit); adjust using the

…†

keys until the DMM reading matches the value shown in DMM
Reading
column of
Table 2-12

Table 2-12. CHANNEL B CC Mode Range I Offset Calibration Parameters

4.

Scroll to next calibration parameter, Item 2 (press

„).

5.

Set DC power supply to +5VDC.

ITEM

MODEL

Load Setting DMM Reading

SLD-60-505-255 0.000

A

0.000A

SLD-61-505-255 0.000

A

0.000A

SLD-60-20-102 0.0000A 0.0000A
SLD-80-20-102 0.0000A

6.

PRES ON (press PRES key until its LED is lit); set load current setting
to match the value shown in Load Setting column of Table 2-13, for
the model being calibrated.

7.

LOAD ON (press LOAD key until its LED is lit); adjust using the

…†

keys until the DMM reading matches the value shown in DMM
Reading
column of Table 2-13, for the model being calibrated.

8.

Repeat Steps 1 through 7 (press

ƒ to return to previous item) until no

further adjustments are necessary for the DMM reading to match the
value in the DMM Reading column for both calibration parameters
(Items 1 and 2).

ITEM

MODEL

Load Setting DMM Reading

SLD-60-505-255 0.500

A

0.500A

SLD-61-505-255 0.500

A

0.500A

SLD-60-20-102 2.002A 2.002A
SLD-80-20-102 2.002A 2.002A
SLD-61-5-752 0.5000A 0.5000A
SLD-62-5-752 0.5000A 0.5000A

2

SLD-60-105-550 0.5000A

0.5000A

Table 2-13. Channel B CC Mode Range 1 Gain Adjustment Parameters

9.

Scroll to next calibration parameter, Item 3 (press

„).

0.0000A

SLD-61-5-752 0.0000A 0.0000A
SLD-62-5-752 0.0000A 0.0000A

1

SLD-60-105-550 0.0000A

0.0000A

SLD Calibration Manual