Calibration Procedure

Elgar Electronics Corporation

Range II Offset and Gain, Channel B CC Mode

1.

Set DC power supply to +5VDC.

2.

PRES ON (press PRES key until its LED is lit); set load current setting
to match the value shown in Load Setting column of Table 2-14, for
the model being calibrated.

3.

LOAD ON (press LOAD key until its LED is lit); adjust using the

…†

keys until the DMM reading matches the value shown in DMM
Reading
column of Table 2-14, for the model being calibrated.

ITEM

MODEL

Load Setting DMM Reading

SLD-60-505-255 0.00

A

0.000A

SLD-61-505-255 0.00

A

0.000A

SLD-60-20-102 0.000A 0.000A

SLD-80-20-102 0.000A 0.000A

SLD-61-5-752 0.000A 0.000A

SLD-62-5-752 0.000A 0.000A

3

SLD-60-105-550 0.000A

0.000A

Table 2-14. CHANNEL B CC Mode Range II Offset Adjust Pparameters.

4.

Scroll to next calibration parameter, Item 4 (press

„).

5.

Set DC power supply to +5VDC at the input terminal.

6.

PRES ON (press PRES key until its LED is lit); set load current setting
to match the value shown in Load Setting column of Table 2-15, for
the model being calibrated.

7.

LOAD ON (press LOAD key until its LED is lit); adjust using

…† keys

until the DMM reading matches the value shown in DMM Reading
column of Table 2-15, for the model being calibrated.

8.

Repeat Steps 1 through 7 (press

ƒ to return to previous item) until no

further adjustments are necessary for the DMM reading to match the
value in the DMM Reading column for both calibration parameters
(Items 3 and 4).

ITEM

MODEL

Load Setting DMM Reading

SLD-60-505-255 5.000

A

5.000A

SLD-61-505-255 5.000

A

5.000A

SLD-60-20-102 20.00

A 20.00A

SLD-80-20-102 20.00

A 20.00A

SLD-61-5-752 5.000

A 5.000A

SLD-62-5-752 5.000

A 5.000A

4

SLD-60-105-550 100.01

A 100.01A

Table 2-15. Channel B CC Mode Range II Gain Adjustment Parameters

9.

Scroll to next calibration parameter, Item 5 (press

„).

2-16

SLD Calibration Manual