Calibration Procedure

CR Range I Offset and Gain, Channel B CR Mode

1.

Set load to match the value in CR Range I Offset column of
Table 2-19, for the model being calibrated.

2.

Set the DC Voltage Standard to match the value in V Sense Input
column of Table 2-19 for the same model being calibrated.

3.

LOAD ON (press LOAD key until its LED is lit).

4.

Adjust using

…† keys until DMM reading matches the value in DMM

Reading column of Table 2-19.

ITEM

MODEL

CR Range I OFFSET

V SENSE input DMM Reading

SLD-60-505-255

4500

60.00 V

0.013 A

SLD-61-505-255

4500

60.00 V

0.013 A

SLD-60-20-102

11250 Ω

60.00 V

0.0053 A

SLD-80-20-102

15000

80.00V 0.0053

A

SLD-61-5-752

45000

60.00 V

0.0013 A

SLD-62-5-752

45000

60.00 V

0.0013 A

15

SLD-60-105-550 45000

60.00 V

0.0013 A

Table 2-19. Channel B CR Range I Offset Calibration Settings

5.

Scroll to next calibration parameter, Item 16 (press

„).

6.

Set the DC Voltage Standard to match the value in IV Sense Input
column of Table 2-20 for the same model being calibrated.

7.

LOAD ON (press LOAD key until its LED is lit). Adjust using

…† keys

until DMM reading matches the value in DMM Reading column of
Table 2-20.

8.

Repeat Steps 1 through 7 (press

ƒ to return to previous item) until no

further adjustments are necessary for the DMM reading to match the
value in the DMM Reading column for both calibration parameters
(Items 15 and 16).

ITEM MODEL CR Range I GAIN V SENSE input

DMM Reading

SLD-60-505-255 1.200

60.00

V 5.00 A

SLD-61-505-255 1.200

60.00

V 5.00 A

SLD-60-20-102 3.000

15.00

V 5.000 A

SLD-80-20-102 4.000

15.00V 3.750 A

SLD-61-5-752 12.000

60.00

V 5.00 A

SLD-62-5-752

12.000

60.00 V

5.00 A

16

SLD-60-105-550

12.000

60.00 V

5.00 A

Table 2-20. Channel B CR Mode Range I Gain Calibration Setting

9.

Scroll to next calibration parameter, Item 17 (press

„).

SLD Calibration Manual