AMETEK SL Series Software User Manual
Page 19

Elgar Electronics Corporation
Calibration Procedure
CR Range I Offset and Gain, Channel A CR Mode
1.
Set load to match the value in CR Range I Offset column of Table 2-8,
for the model being calibrated.
2.
Set the DC Voltage Standard to match the value in IV Sense Input
column of Table 2-8 for the same model being calibrated.
3.
LOAD ON (press LOAD key until its LED is lit).
4.
Adjust using
keys until DMM reading matches the value in DMM
Reading column of Table 2-8.
ITEM MODEL
CR Range I OFFSET V SENSE input
DMM Reading
SLD-60-505-255
4500
Ω
60.00 V
0.013 A
SLD-61-505-255
4500
Ω
60.00 V
0.013 A
SLD-60-20-102
11250
Ω
60.00 V
0.0053 A
SLD-80-20-102
15000
Ω
80.00V 0.0053
A
SLD-61-5-752
45000
Ω
60.00 V
0.0013 A
SLD-62-5-752
45000
Ω
60.00 V
0.0013 A
15
SLD-60-105-550
2250
Ω
60.00 V
0.026 A
Table 2-8. Channel A CR Mode Range I Offset Calibration Settings
5.
Scroll to next calibration parameter, Item 16 (press
).
6.
Set the DC Voltage Standard to match the value in V Sense Input
column of Table 2-9 for the same model being calibrated.
7.
LOAD ON (press LOAD key until its LED is lit). Adjust using
keys
until DMM reading matches the value in DMM Reading column of
Table 2-9.
8.
Repeat Steps 1 through 7 (press
to return to previous item) until no
further adjustments are necessary for the DMM reading to match the
value in the DMM Reading column for both calibration parameters
(Items 15 and 16).
ITEM MODEL CR Range I GAIN V SENSE input
DMM Reading
SLD-60-505-255
1.200
Ω
15.00 V
12.50 A
SLD-61-505-255
1.200
Ω
15.00 V
12.50 A
SLD-60-20-102
3.000
Ω
15.00 V
5.000 A
SLD-80-20-102
4.000
Ω
15.00V 3.750
A
SLD-61-5-752
12.000
Ω
30.00 V
2.500 A
SLD-62-5-752
12.000
Ω
30.00 V
2.500 A
16
SLD-60-105-550
0.600
Ω
6.000 V
10.00 A
Table 2-9. Channel A CR Mode Range I Gain Calibration Setting
SLD Calibration Manual
2-11