beautypg.com

Table 2-5. channel a dvm calibration settings -6 – AMETEK SL Series Software User Manual

Page 14

background image

Calibration Procedure

Elgar Electronics Corporation

DVM, Channel A

1.

Set the DC Voltage Standard to the value shown in DC Input column
of Table 2-5.

2.

Adjust using

…† keys until the DMM reading matches the value in

DMM Reading column.

3.

Scroll to next item (press

„) and repeat Steps 1 and 2 for each item

through Item 8.

4.

Repeat Steps 1 through 3 for all calibration parameters, Items 5
through 8 (press

ƒ three times to return to Item 5) until no further

adjustments are necessary for the DMM reading to match the value in
the DMM Reading column of Table 2-5.

ITEM

DC INPUT

DMM Reading

5

0.000 V

0.000 V

6

15.0000 V

15.000 V

7

0.000 V

0.000 V

8

60.000 V

60.00 V

*

80.000 V

80.00 V

*

Applies to SLD-80-20-102 only

Table 2-5. Channel A DVM Calibration Settings

5.

Scroll to next calibration parameter, Item 9 (press

„).

2-6

SLD Calibration Manual