Verification and Calibration

Sorensen SFA Series

4.3.1

C

URRENT

M

ODE

1. Set the SFA Series unit to operate in remote current programming

mode using an external 0-5 Vdc voltage source as shown in Figure 3–6
in Section 3.3 Remote Current Programming.

2. Attach a precision meter across the shunt Kelvin terminals.

3. Attach a precision meter in parallel with the voltage programming

source.

4. Set the voltage source to 0.0V

±

1mV.

5. Apply AC power and verify the current mode indicator is illuminated.

6. Verify that the unit is set to 0A

±

0.8% of full-scale output. If necessary,

adjust R47 for zero current through the shunt (see Figure 4-1 in Section
4.1.2 Preparation).

7. Set the programming voltage to 5.0V

±

1mV.

8. Verify that the unit is set to 100%

±

0.8% of full scale output current. If

necessary, adjust R33 for 100% of full scale current on the shunt.

9. Set the programming source for 0.5V

±

1mV.

10. Verify that the unit is set to 10%

±

0.8% of full scale output current. If

necessary, adjust R47 for 10% of full scale current on the shunt.

11. Repeat the steps above as required to obtain the required accuracy.

12. Set the programming source to obtain 100% output current on the

shunt. Verify that the front panel meter displays 100%

±

1% of full-

scale output current. If necessary adjust R85 such that the current
display reads 100% current.

CAUTION!

Hazardous voltages may be present on the output after the output is disabled.
Allow 5 minutes to drain any charge to safe levels before connecting or removing
output wiring.