Installation and Configuration

Basic Setup Procedure


Basic Setup Procedure

This procedure can be used as a quick reference for those familiar with the
configuration requirements for the GPIB-M interface as installed in the DC power
supply.

IEEE-488 Controller Connection

Connect the GPIB cable to the supply at connector on the rear panel. Use an approved
IEEE-488 connector and cable when connecting the GPIB Interface to your IEEE-488
GPIB network. Refer to

Figure 2.4

,

Figure 2.5

,

Figure 2.6

or

Figure 2.7

.

Add a ferrite block to reduce radiated emissions. The one inch square ferrite block
with built-in housing clip is packaged and shipped with the power supply interface
card.

To install the ferrite block:

1.

Position the block no more than 5 cm (2 in.) from the power supply end of the
user cable.

2.

Open the ferrite block housing.

3.

Run the cable through the ferrite block. See

Figure 2.9, “XFR GPIB Cable with

Ferrite Block” on page 31

.

4.

Close the housing clip.

The ferrite block ensures that the power supply system meets radiated emission
requirements for CE Mark.

Figure 2.9 XFR GPIB Cable with Ferrite Block