AMETEK HPD Series GPIB-Multichannel User Manual
Page 111

SCPI Com
m
an
d Refer
e
nce
SCPI Co
mmand Summary
Release A
10
7
Table B.4Commands for Current Share
Table B.5Commands for Calibration
Table B.6Command to Clear all Protection Mechanisms
Set High Voltage Limit
[[:]SOURce][
{
Set upper limit of voltage setpoint range (soft
limits)
Yes
Set Low Voltage Limit
[[:]SOURce][
{
Set lower limit of voltage setpoint range (soft
limits)
Yes
Function
SCPI Command
Description
Query
Set Current Sharing Mode
[[:]SOURce][
{NONE|MASTer|SLAVe}
Select current share mode
Yes
Read Summed Current
[:]MEASure[:SCALar]:CURRent[:DC]? SUM
Read total current output of all current
sharing supplies
N/A
Function
SCPI Command
Description
Query
Restore Factory
ion
[:]CALibration[
Restonres the calibration to the constants set
at the factory
N/A
Change Calibration
Password
[:]CALibration[
Changes the calibration security code.
No
Set Calibration State
[:]CALibration[
Change calibration state (mode)
Yes
Set Supply Output Current
Level
[:]CALibration[
{MINimum|MAXimum}
Set output current calibration level
No
Enter Output Current Data
[:]CALibration[
Set output current calibration data
No
Set Supply Output Voltage
Level
[:]CALibration[
{MINimum|MAXimum}
Set voltage output calibration level
No
Enter Output Voltage Data
[:]CALibration[
Set voltage output calibration data
No
Function
SCPI Command
Description
Query
Clear Output Protection
[:]OUTPut[
Clears the protection mechanism.
N/A