Expressions – AMETEK HPD Series GPIB-Multichannel User Manual
Page 119

SCPI Com
m
an
d Refer
e
nce
Expressions
Release A
11
5
Expressions
Table B.15Expressions
Expression
Details
aux_line_mnemonic
Define the output of the auxiliary line.The possible values are NONE, ON, OFF, OVOLtage,
UVOLtage, OCURrent, UCURrent, ACOFf, OTEMperature, SPRotection, UNRegulated, FOLD, CC,
CV, CP.
channel
The address for a multichannel slave. An integer value in the range 2 to 50.
codeword
A string representing any 4-digit positive integer.
current
A numeric value as defined by SCPI. May also be MAXimum or MINimum. May include
current-related suffix units such as “mA”, “uA”, “A” etc. Range may be 0 to 103% of model’s rated
current.
delay
A length of time in the range 0 to 60 seconds. May include time-related suffix units such as “S”,
“MIN”, “mS”, “uS” etc. By default, the value is in seconds.
ESE-word
Range 0–255. An 8-bit status mask for the Standard Event Status Register that determines which
bits are OR’d to form the ESB bit in the Status Byte Register.
GPIB_address
The address for a GPIB controlled unit. An integer value in the range 1 to 30.
on-off-state
A Boolean indicator of a state. Possible values are ON, OFF, 0 (off), or 1 (on).
OVP_level
A numeric value as defined by SCPI. May also be MAXimum or MINimum. May include
voltage-related suffix units such as “mV”, “uV”, “V”, etc. Range is 0 to 103% of model’s rated voltage.
setting_location
A numeric indication of an internal set of setting registers. An integer value in the range 1 to 10.
sequence_count
Number of times a sequence is to be repeated. Range is 1 to 9999.
sequence_number
The name of an auto sequence program. Range is 1 to 10. The suffix is part of the SEQUENCE
command name and is not a parameter.
status-enable
A 16-bit status mask for any condition register that determines which bits are to be used for
synthesizing the summary bit of that register.
step_number
The step number of an auto sequence program. Possibly considered a SCPI suffix. Range is 1 to
99.
step_time
The duration of an auto sequence step in the format hh:mm:ss.s.
May include time-related suffix units such as “S”, “MIN”, “mS”, “uS”, etc. By default, the value is in
seconds.
voltage
A numeric value as defined by SCPI. May also be MAXimum or MINimum. May include
voltage-related suffix units such as “mV”, “uV”, “V”, etc. Range is 0 to 103% of model’s rated voltage.