beautypg.com

Current share error codes – AMETEK HPD Series GPIB-Multichannel User Manual

Page 128

background image

Error Messages
Current Share Error Codes

124

Operating Manual for Multichannel Functionality (GPIB-M)

Current Share Error Codes

Table C.13Current Share Error Codes

Error
code

Error Message Description

+1900

Current Share General Error

+1911

Current share master already online
A unit on the CANbus network has already been assigned the current share
master unit. Only one is allowed per network.

+1912

Current share slave lost
One of the connected current share slaves have not responded in time.

+1921

Current share master lost
The assigned master has not responded in time.

+1922

Current share slave connection refused
Connection as a slave on the current share network has been refused due to
one of the following reasons:
4 slaves are already online,
the voltage rating does not match those of the master’s,
the current rating does not match those of the master’s.

+1924

Current share data out of range
The voltage or current readback or setpoint values passed are out of range.

+1925

Current share slave output off by unknown
One of the slave’s output is off because of unknown reasons

+1926

Current share slave output off by command
One of the slave’s output is off because of a command from a remote source
or the front panel.

+1927

Current share slave output off by AC fail
One of the slave’s output is off because of an AC fail condition.

+1928

Current share slave output off by OTP (Over Temperature Protection)
One of the slave’s output is off because of an OTP condition.