beautypg.com

2 description – Casella CEL Microtherm WBGT heat stress monitor User Manual

Page 8

background image

Table 1: Permissible Heat Exposure Threshold Limit Values

(values are given in

o

C WBGT)

Work Regimen

Work Load

Light

Moderate

Heavy

Continuous Work

30.0

27.7

25.0

75% Work

30.6

28.0

25.9

25% Rest, Each Hour
50% Work

31.4

29.4

27.9

50% Rest, Each Hour
25% Work

32.2

31.1

30.0

75% Rest, Each Hour

1.2

Description

The Casella Microtherm Heat Stress Wet Bulb Globe Temperature (WBGT)

monitor is shown in Figure 1.

Knurled
locking ring

Radiation shield

with dry sensor

inside

Globe temperature
sensor

Wet temperature

sensor

Water reservoir

Cotton wick

Sensor array

Control unit

Escape

Navigator keys

Enter

Battery charging

indicator

ON / OFF

00061

MICROTHERM

Heat Stress WBGT

Heat Stress WBGT

Figure 1: General layout of the Microtherm Heat Stress WBGT

Page 8 of 44

Description

MICROTHERM Heat Stress WBGT &

WinHSM Application Software - User Manual

This manual is related to the following products: