Additional Displayed Information
Battery voltage,
Charger circuit status,
Memory status,
Number of runs saved,
Firmware version,
User name and details.
Data Logger
Internal memory:
512 k RAM (with back-up battery) that
provides space for over 49 000 intervals,
so that when all five temperature and
humidity parameters are enabled, a total
of 245 000 individual results can be saved.
Logger Interval:
Adjustable from 30 seconds to 60 minutes.
Recorded Values:
Measured levels for the logging interval.
Maximum, minimum and average values
for the entire logging period are saved at
the end of each logging run.
Serial Interface:
RS232 up to 38,400 baud.
Infrared Interface (Future expansion).
Alarm output:
+5 V DC when the alarm is activated,
0 V while the alarm is inactive.
5.2
WinHSM, PC Based Application Software
Provides a friendly graphical interface for instrument configuration, data
download and graphical report presentation.
Operating System
M.S. Windows 95/98/ME/2000 and
NT compatible,
Pentium II or better processor,
200 MHz recommended for real time
mode,
Requires 1.44 MB 3
1
/
2
" floppy disk drive,
RS232 serial port,
16 MB RAM, FDD and HDD,
Super VGA colour graphics.
5.3
Spare Parts
107039A
Spare Wicks (pack of 10).