beautypg.com

Casella CEL Microtherm WBGT heat stress monitor User Manual

Page 41

background image

5.

SPECIFICATION

5.1

Instrument Specification

Measurement

Sensing technique:

Platinum Resistance Temperature

Detectors, 100 Ω elements to EN 60751

Class A using 4-wire cable up to 30 m,

Ta, Dry Bulb (Air):

Measuring Range: 10 - 60

o

C,

Accuracy: ±0.5

o

C,

Tg, Globe:

Measuring Range: 20 - 120

o

C,

Accuracy 20 - 50

o

C: ±0.5

o

C,

Accuracy 50 - 120

o

C: ±1.0

o

C,

Tnw, Natural Wet:

Measuring Range: 5 - 40

o

C,

Accuracy: ±0.5

o

C,

Relative Humidity:

Less than ±5% from 35 - 100% RH,

Resolution:

0.1

o

for temperature,

0.1% for RH.

For inside/outside measurements with no solar load:

Use WBGTin = 0.7Tnw + 0.3Ta.

For outside measurements with solar load:

Use WBGTout = 0.7Tnw + 0.2Tg + 0.1Ta.

Power Supply

Instrument Consumption: Approx. 30 mA.

Battery:

4 x AA / MN1500 cells Alkaline or

rechargeable Nickel Cadmium.

Typical operating time:

Alkaline Batteries (2700 mAh), >85 hours,

Rechargeable NiCd cells (950 mAh),

>30 hours

Battery Charging

Internal NiCd fast charger circuitry.

Charge termination based on Negative

delta V detection with additional time-out

protection.

The charger must be disabled for dry cell

operation.

Charge Rate

Fast charge rate 450 mA,

Standby charge rate 55 mA approx.

Power adapter

Universal with input voltage range

100 - 240 V AC, 47 - 63 Hz.

Output 12 V DC at 0.8 A.

Supplied with connectors for UK, USA

and European mains outlet sockets. The

instrument may be powered continuously

from the external power adapter.

Page 41 of 44

Specification

MICROTHERM Heat Stress WBGT &
WinHSM Application Software - User Manual