beautypg.com

Casella CEL Microtherm WBGT heat stress monitor User Manual

Page 39

background image

4.

SERVICING

4.1

Service Department

Casella CEL's in house service department offers a comprehensive range of

repair and calibration services, designed to effect a fast and efficient back-up

for all our products. The Service Department is operated under the scope of

our BSI registration for products manufactured by us. We will however,

undertake the repair of other manufacturers equipment.
For further information please contact our service department at our Bedford

headquarters. We will be happy to provide quotations for individual repairs or

provide annual maintenance under contract.
We recommend factory service by technicians trained and equipped to repair

your instrumentation. Should you wish factory repair assistance, send your

equipment in a package equivalent to the original packaging. Insure to full

value and ship pre-paid. Include a letter giving full details with your packing

list.

Send to:

Casella CEL Ltd

(Service Department)

Regent House

Wolseley Road

Kempston

Beds MK42 7JY

United Kingdom

For service outside the United Kingdom, please return to our appointed

distributor.

4.2

Fault Finding Tips

The Microtherm Heat Stress WBGT contains no user serviceable

components. Repairs should be performed only by authorised personnel.

Fault Symptom

Possible Cause

Service Hint

Instrument does not

switch ON:

Batteries discharged, or

failing to charge. Confirm

charger indicator is lit.

Recharge or replace

batteries if necessary.

Use external power supply.

Rechargeable batteries not

charging:

Charging light not lit:

Faulty rechargeable battery

or batteries.

Faulty power supply.

Battery charger switch

disabled.

Check batteries and

connections.

Replace power supply.

Enable charger switch (see

Section 2.1.1).

Sensor Error - message

shown on instrument

display:

Missing or damaged

sensor.

Damaged connector.

Check condition of sensor

and that it is correctly fitted.

Page 39 of 44