Casella CEL Microtherm WBGT heat stress monitor User Manual
Page 23

The logger interval is the time between each recorded data point
The following options are available:
30 seconds,
1, 2, 3, 4, 5, 6, 10, 15, 20, 30, 60 minutes.
4.
If necessary, change the current setting.
5.
Select logger Mode.
The logger mode determines what happens when all empty memory
locations have been filled with data.
6.
If necessary, change the setting to one of the following alternatives:
Over Write Run
The logger simply carries on recording
data so that the oldest records will be
overwritten by new ones.
Stop When Full
The logger stops logging when the
memory is full.
2.5.2 Clearing the Logger Memory
The Microtherm has a logging capacity of over 40 000 intervals (giving more
than 245 000 individual temperature or relative humidity results) and 32 runs,
with the total number of stored runs shown on the Status screen.
When there is insufficient space remaining in the instrument store for the
projected logging run, stored data can be deleted. All recorded data should be
downloaded before clearing the logger's memory as there are no facilities to
recover data deleted from the instrument.
1.
Select and display the Config Logger screen (see Section 2.5.1).
2.
Highlight the Clear Memory option, select YES then press
.
A warning message will be displayed asking for confirmation before
the contents of the memory are erased.
3.
Press
to confirm the erase operation.
ALL stored runs will be DELETED.
2.5.3 Start Logging
1.
With the logger settings configured as described in Section 2.5.1,
(and the memory cleared when necessary) display the Logger screen.
Enter
Enter
Page 23 of 44
Operation
MICROTHERM Heat Stress WBGT &
WinHSM Application Software - User Manual