BINDER KBF 115 User Manual
Page 96

KBF / KBF P (E5.3) 09/2014
page 96/110
Unit size
115
240
720
Electrical data
IP system of protection acc. to EN 60529
20
20
20
Nominal voltage (+/-10%) 50/60 Hz
V
200 to 240 1N~
Power plug
shock proof plug
Nominal power
kW
2.00
2.10
3.10
Energy consumption 4) at 40 °C / 104 °F and 75 %
r.H.
Wh/h
470
650
620
Installation category acc. to IEC 61010-1
II
II
II
Pollution degree acc. to IEC 61010-1
2
2
2
Over-current release category B, 2 poles
Amp
16
16
16
Noise level approx.
dB (A)
52
52
53
Electrical connection data KBF constructed acc. to cUL requirements (for the USA and Canada)
Unit size
115
240
720
Electrical data KBF-UL
IP system of protection acc. to EN 60529
20
20
20
Nominal voltage (+/-10%) 60 Hz
V
200 to 240 / 2~
Nominal power
kW
2.00
2.10
3.10
Power plug
NEMA
6-20P
6-20P
L6-20P
Installation category acc. to IEC 61010-1
II
II
II
Pollution degree acc. to IEC 61010-1
2
2
2
Over-current release category B, 2 poles
Amp
16
16
16
1)
Lower values are valid at an ambient temperature of max. 25 °C / 77 °F
2)
to 98 % of the set value
3)
Use this value for sizing air condition systems
All technical data is specified for unloaded units with standard equipment at an ambient temperature of
+25 °C / 77 °F and a power supply voltage fluctuation of +/-10%. The temperature data is determined in
accordance to BINDER factory standard following DIN 12880, observing the recommended wall clear-
ances of 10 % of the height, width and depth of the inner chamber. Technical data refers to 100% fan
speed.
All indications are average values, typical for units produced in series. We reserve the right to
change technical specifications at any time.
If the chamber is fully loaded, the specified heating up and cooling down times may vary ac-
cording to the load.
Bringing a source of humidity into the inner chamber will affect the minimum humidity specifi-
cation and may affect the humidity accuracy.