Caution – BINDER KBF 115 User Manual
Page 13

KBF / KBF P (E5.3) 09/2014
page 13/110
Indications on the type plate (example) Information
50/60 Hz
Power frequency
Max. operating pressure 15 bar
Max operating pressure in the refrigerating system
(15 bar / 218 PSI)
R 134a – 0,17 kg
Refrigerant type and filling weight (0.17 kg / 0.37 lbs)
Contains fluorinated greenhouse gases covered by the Kyoto Protocol
Symbol on the type plate
Information
CE coformity marking
Electrical and electronic equipment manufactured / placed
on the market in the EC after 13 August 2005 and be dis-
posed of in separate collection according to the directive
2002/96/EC on waste electrical and electronic equipment
(WEEE).
The equipment is certified in the GOST R certification sys-
tem of GOSTSTANDARD Russia.
LABORATORY EQUIPMENT
43KM
(UL units only)
The equipment is certified by Underwriters Laboratories
Inc.
®
according to standards CAN/CSA-C22.2 No. 61010-1,
2
nd
Edition, 2004-07 (Electrical Equipment for Measure-
ment, Control, and Laboratory Use; Part 1: General Re-
quirements); UL 61010-1, 2
nd
Edition, 2005-07-22 (Electri-
cal Equipment for Measurement, Control, and Laboratory
Use; Part 1: General Requirements); IEC 61010-1:2001, 2
nd
Edition and IEC 61010-2-10 (Particular Requirements for
Laboratory Equipment for the heating of materials).
1.5 General safety instructions on installing and operating the constant climate
chamber
With regard to operating the constant climate chamber KBF / KBF P and to the installation location,
please observe the guideline BGI/GUV-I 850-0 on safe working in laboratories (formerly BGR/GUV-R 120
or ZH 1/119 laboratory guidelines issued by the employers’ liability insurance association) (for Germany.
BINDER GmbH is only responsible for the safety features of the unit provided skilled electricians or quali-
fied personnel authorized by BINDER perform all maintenance and repair, and if components relating to
chamber safety are replaced in the event of failure with original spare parts.
To operate the unit, use only original BINDER accessories or accessories from third-party suppliers au-
thorized by BINDER. The user is responsible for any risk caused by using unauthorized accessories.
CAUTION
Danger of overheating.
Damage to the unit.
∅
Do NOT install the unit in unventilated recesses.
Ensure sufficient ventilation for dispersal of the heat.