6 intended use, Warning, Danger – BINDER KBF 115 User Manual
Page 15

KBF / KBF P (E5.3) 09/2014
page 15/110
WARNING
Stability hazard.
Danger of injury.
Damage to the unit and the charging material.
Housing cover breakaway.
∅
Do NOT climb on the lower housing cover.
∅
Do NOT load the lower housing cover with heavy objects while the unit door is open.
1.6 Intended use
Constant climate chambers series KBF / KBF P are suitable for exact conditioning of harmless materials.
A mixture of any component of the charging material with air must NOT be explosive. The operating tem-
perature must lie below the flash point or below the sublimation point of the charging material. Any com-
ponent of the charging material must NOT be able to release toxic gases.
Other applications are not approved.
Constant climate chambers series KBF / KBF P are not classified as medical devices as defined
by the Medical Device Directive 93/42/EEC.
Following the instructions in this operating manual and conducting regular maintenance work
(chap. 16) are part of the intended use.
DANGER
Explosion or implosion hazard.
Danger of poisoning.
Danger of death.
∅
Do NOT introduce any substance combustible or explosive at working temperature into
the constant climate chamber.
∅
NO explosive dust or air-solvent mixture in the inner chamber.
∅
Do NOT introduce any substance which could lead to release of toxic gases.
The charging material shall not contain any corrosive ingredients that may damage the ma-
chine components made of stainless steel, aluminum, and copper. Such ingredients include in
particular acids and halides. Any corrosive damage caused by such ingredients is excluded
from liability by BINDER GmbH.
WARNING: If customer should use a BINDER chamber running in non-supervised continuous
operation, we strongly recommend in case of inclusion of irrecoverable specimen or samples
to split such specimen or samples and store them in at least two chambers, if this is feasible.