page 26/110
4. Installation and connections
4.1 Spacers for rear wall distance
Please fix both spacers with the supplied screws at the unit rear. This serves to ensure the prescribed
minimum distance to the rear wall of 100 mm / 3.94 in.
Figure 10: Spacer for rear wall distance
Figure 11: Rear KBF / KBF P with mounted spacers
4.2 Wastewater connection
Fasten the wastewater hose to the wastewater connection “OUT” (17) on the rear of the unit (olive ∅ 14
mm). Observe the following points:
•
You can use a part of the supplied water hose as a drainage hose. In case another hose is used, it
has to be permanently resistant against at least 95 °C / 203 °F.
•
Mount the wastewater hose with a maximum positive inclination of 1 m / 3.28 ft and a maximum total
length of 3 m / 9.8 ft.
•
Protect both ends of the drainage hose with two of the four supplied hose clamps.
Wastewater is collected in an internal can with a volume of approx. 0.5 liters. It is pumped off
only when required, thus there is no continuous wastewater flow.
Protect the wastewater supply at both sides with the supplied hose clamps.