3 performance after power failure in manual mode, Program operation – BINDER KBF 115 User Manual
Page 45

KBF / KBF P (E5.3) 09/2014
page 45/110
8.2 Entering the set point values for KBF P when operating without illumination
The KBF P has been adjusted for operation with illumination. Since the illumination creates a heat input in
the chamber, this must be considered when entering a setpoint without illumination. For this reason,
adapting the setting of the temperature set-point according to the following table is required.
Target temperature value
Set-point entry without illumination
KBF P 240
KBF P 720
10 °C
10 °C
11 °C
20 °C
21 °C
21 °C
30 °C
31 °C
32 °C
40 °C
41 °C
42 °C
50 °C
51 °C
53 °C
60 °C
62 °C
63 °C
70 °C
72 °C
74 C
Adapting the setting of the humidity set-point according to the following table is also required.
Target humidity value
Set-point entry without illumination
KBF P 240
KBF P 720
10 % r.H.
10 % r.H.
9 % r.H.
20 % r.H.
19 % r.H.
17 % r.H.
30 % r.H.
26 % r.H.
25 % r.H.
40 % r.H.
36 % r.H.
34 % r.H.
50 % r.H.
45 % r.H.
43 % r.H.
60 % r.H.
55 % r.H.
52 % r.H.
70 % r.H.
65 % r.H.
61 % r.H.
80 % r.H.
75 % r.H.
71 % r.H.
8.3 Performance after power failure in Manual Mode
In Manual Mode (HAND), all functions return exactly to the same status the chamber had before power
failure. The set-points are immediately resumed. No error message indicating that a power failure has
taken place is displayed. However, the power failure will appear in the event list.
9.
Program operation
The MB1 2-channel program controller permits programming temperature and humidity cycles. It offers
25 program memory positions with 100 program sections each. The total cumulative number of program
sections is limited to 500. It is not possible to link several programs.
For each program section a temperature set-point, a humidity set-point, the fan speed (0% to 100%), and
the switching-state of up to 8 operation lines (non-functional with standard unit) can be entered.
Programming is possible directly by the keypad of the controller or graphically by the software APT-
COM™ 3 DataControlSystem (option, chap. 15.1) specially developed by BINDER.