KBF / KBF P (E5.3) 09/2014
page 80/110
(2) Cable connections
Connect the plug of the cable to the socket (22) at the rear of the unit.
The socket (22) is marked with a sticker:
Socket (22)
(17)
(18)
Figure 31: Connections at the unit rear KBF / KBF P 240
(3) Hose connections
Plug the freshwater hose into the hose connection (21) above the freshwater can and secure it with
a hose clamp. You can use a part of the standard supplied water hose.
Screw the hose nozzle (brass) to the free edge of the hose and screw it directly onto the freshwa-
ter connection “IN” (18) at the rear of the unit.
When the freshwater can is empty, within 60 seconds the alarm message “HUMID ALARM” will be dis-
played on the controller, the buzzer sounds (chap. 11), and the humidity system turns off.
To guarantee humidification during 24 hours even at high humidity set-points with manual
water supply, we recommend filling the freshwater can (option) at the end of each day.