CalMAN Meter Connection Procedure
1. On the Meter Settings tab, click Find Meter.
2. On the Find Meters popup dialog:
a. Leave Com Port set to auto, or, select the com port that was assigned to the meter
(shown in Device Manager, under Ports (com &LPT)).
b. Check the box for “Colorimetry Research Inc Meters (USB).”
c. Click Search.
3. On the Meter Settings tab:
a. Select the calibration matrix for your target display type.
b. Select the Sync Mode that produces the most consistent measurements on the target
display (check with Read Continuous on 100% White).
i. Off: The default sync mode for the CRI meter. This is the fastest mode of meter
operation, but may not provide consistent measurements on some display
types.
ii. On: Takes one sync measurement to synchronize the meter with the display
refresh rate, then performs all of the series or continuous data measurements.
This is only slightly slower, but may not provide optimum sync with displays that
have different refresh rates at lower luminance levels.
iii. Auto: Takes a sync measurement prior to each data measurement. This is
slower yet, but provides optimum sync with all display types, at all luminance
levels.