beautypg.com

SOR 815PT Smart Pressure Transmitter User Manual

Page 12

background image

12/24

Form 1571 (10.13) ©SOR Inc.

HART

Command

Description

Command Data

Command Reply Data

Byte

Description

Byte

Description

18

Write tag,

descriptor, date

0-5

Tag

0-5

Tag

6-17

Descriptor

6-17

Descriptor

18-20

Date

(dd-mm-yyyy)

18-20

Date (dd-mm-yyyy)

19

Write final

assembly

number

0-2

Final assembly

number

0-2

Final assembly number

20

Read long tag

Read only

0-31

Long tag

21

Read unique

identifier

associated with

long tag

0-31

Long tag

0

Always 254

1

Manufacturer ID

2

Device type code

3

Number of preambles

4

Universal command revision

5

Transmit-specific command

revision

6

Software revision

7

Hardware revision

8

Device function flags

9-11

Device ID number

12

Minimum number of preambles

13

Maximum number of preambles

14-15

Configuration change counter

16

Extended field device status

22

Write long tag

0-31

Long tag

0-31

Long tag

33

Read device

variables

0

Transmit variable

code slot 0

0

Transmit variable code slot 0

1

Transmit variable

code slot 1

1

Units code slot 0

2

Transmit variable

code slot 2

2-5

Variable slot 0

3

Transmit variable

code slot 3

6

Transmit variable code slot 1

7

Units code slot 1

8-11

Variable slot 1

12

Transmit variable code slot 2

13

Units code slot 2

14-17

Variable slot 2

18

Transmit variable code slot 3

19

Units code slot 3

20-23

Variable slot 3