Menu descriptions – Parr Instrument 6755 User Manual
Page 18
4
Menu Descriptions
P a r r I n s t r u m e n t C o m p a n y
16
Data Entry Controls Menu
Auto Sample ID Controls: Accesses sub-menu for
controlling the automatic assignment of sample
identifi cation numbers.
Auto Sample ID Controls
Automatic Sample ID: When set to ON the unit
will automatically assign sample identifi cation
numbers in accordance with parameters set by
the other three keys on this menu. When set
to OFF, the user manually enters each sample
ID when prompted to do so.
Auto Sample ID Prefi x: An entry here will be
used as a prefi x for all sample IDs, if the Auto-
matic Sample ID is set to ON. Press this key
to access a sub-menu for entering an alphanu-
meric prefi x.
Next Auto Sample ID Number: Establishes the
initial sample number for a series of tests and
then shows the next sample ID which will be
assigned. Used when the Automatic Sample
ID is set to ON. Press this key to access a sub-
menu for entering a numeric increment.
Auto Sample ID Increment: Establishes the in-
crement between sample numbers; used when
the Automatic Sample ID is set to ON. Press
this key to access a sub-menu for entering a
numeric increment.
Reporting Controls Menu
Report Width: Toggle this key to set the column
width of the printer to either 40 or 80 columns. Se-
lect 40 when the 1758 Printer is used.
Automatic Reporting: Toggles the automatic report-
ing ON/OFF. When ON, reports will be generated at
the conclusion of the test. When OFF, reports will
only be generated through the Report Menu.
Automatic Report Destination: Toggles to direct the
reports to the Printer port or the screen display.
Individual Printed Reports: When set to ON, will
generate extra line feeds for each report printed. In
the OFF position, only one header will be printed for
a series of tests.