Freeze-dryer EPSILON 1-4 LSCplus
Freeze-dryer EPSILON 2-4 LSCplus

7 Malfunctions and error correction

Version 04/2014, Rev. 1.4 of 09/12/2014 • sb

89

Translation of the original operating manual

Pos: 180 /20 0 Ch rist/3 61 G T-BA L abo r-Pilot (PROJEKTE)/Epsilo n Sta nda rd/0 70 Stö run gen und Fe hler such e/07 0-0 010 -00 31 Eisko nde nsat or- und Stellfläc hent emp era tur E psilon @ 25\ mo d_1 4049 846 659 83_ 68.d ocx @ 185 861 @ 3 @ 1

7.1.3 Insufficient ice condenser temperature

CAUTION

Ensure sufficient ventilation. Do not place any paper, cloth, or similar
material behind or under the unit, since otherwise the air circulation will be
impaired.


The refrigeration unit is equipped with a protective device against
overpressure in the refrigeration system and with a thermal motor
protection switch.
The protective devices trip
• when the ambient temperature is too high
• when the air circulation of the heat exchanger of the refrigeration

system is insufficient

• when the refrigeration system is overloaded.
In these cases, the refrigeration unit will be switched off automatically. If the
permissible operating conditions are re-established after a cool-down
phase of several minutes, the refrigeration unit will be switched on again
automatically.
The malfunctions are displayed in the process and equipment information
window.
The minimum ice condenser temperature of approx. –85°C is reached
when the ice condenser is not loaded and the ice condenser chamber is
evacuated.

Pos: 181 /01 0 Univ ersal mod ule/L eerz eile @ 0\m od_ 120 211 624 450 0_0. docx @ 11 4 @ @ 1

Pos: 182 /20 0 Ch rist/3 60 G T-BA L abo r-Pilot (STANDARDMODU LE)/0 70 Stö run gen und Fe hlers uche 070 -00 21 P rozes s un d Anlag en meldu nge n LSCpl us @ 2 5\m od_ 140 498 406 7666 _68 .docx @ 1 842 65 @ 2 @ 1

7.2 Process and error messages

The LSCplus control system displays the complete process and error
messages, which is why they are not included in this operating manual.
You can order these documents from our service department.