Freeze-dryer EPSILON 1-4 LSCplus
Freeze-dryer EPSILON 2-4 LSCplus
6 Operation
56
Version 04/2014, Rev. 1.4 of 09/12/2014 • sb
Translation of the original operating manual
Network (2)
This dialog box is used to change various system settings.
NOTE
The modifications will not become effective until after a restart of the unit.
Fig. 42: Dialog box "Network"
System properties (3)
This dialog box is used to change various system settings.
Fig. 43: Dialog box "System properties"
Beeper: The beeper sounds in the event of a malfunction, for example.
• If the setting is "On", the beeper sounds every few seconds until the
user acknowledges the message.
• If the setting is "Silent", the beeper sounds once when the malfunction
occurs.
• If the setting is "Off", the beeper will not sound at all.