8 procedures in the event of hazards and accidents, 9 remaining hazards – Martin Christ Epsilon 2-4 LSCplus User Manual
Page 30

Freeze-dryer EPSILON 1-4 LSCplus
Freeze-dryer EPSILON 2-4 LSCplus
3 Safety
30
Version 04/2014, Rev. 1.4 of 09/12/2014 • sb
Translation of the original operating manual
Pos: 75 / 200 Christ /36 0 GT -BA La bor -Pilot (S TANDARDMODULE) /03 0 Sicher heit/ 030 -00 82 V erh alten bei G efah re n un d Unf ällen Epsil on mit H2O 2 @ 2 5\m od_ 140 498 331 447 6_68 .docx @ 1 834 39 @ 2 @ 1
3.8
Procedures in the event of hazards and accidents
DANGER
Fire:
• A fire in the electrical control system must be extinguished with a CO
2
fire extinguisher!
• Burning oil must be extinguished with a CO
2
fire extinguisher or powder
fire extinguisher!
Hazardous electrical incident:
• Set the main switch and the mains power switch of the control system
to the "OFF" or "0" position in order to interrupt the power supply
completely.
Injuries caused by liquids under higher pressure (e.g. hydraulic oil):
• Liquids that leak out under high pressure, e.g. hydraulic oil, can
penetrate the skin and cause severe injuries.
Seek medical treatment immediately! There is a risk of embolism!
Burns:
• Small-area burns (e.g. finger) should be cooled immediately with
lukewarm water for approximately 2 minutes.
• Large-area burns should not be cooled because of the resulting risk of
hypothermia.
• Cover the burns loosely and in a sterile manner (e.g. with a sterile
dressing).
• Cover the injured person with a blanket.
Chemical burns (e.g. caused by H
2
O
2
):
• Eyes:
Rinse for at least 10 minutes with sterilised water.
• Lungs: Remove the injured person immediately to fresh air and keep
them at rest and warm.
• Skin:
Soak with water, remove clothing.
• Mouth: Rinse thoroughly with water, provide plenty to drink.
IF IN DOUBT, CALL THE EMERGENCY PHYSICIAN (AMBULANCE)!
Pos: 76 / 010 Unive rsalm odul e/Le erzeil e @ 0\mo d_1 202 116 244 500 _0.d ocx @ 114 @ @ 1
Pos: 77 / 200 Christ /36 0 GT -BA La bor -Pilot (S TANDARDMODULE) /03 0 Sicher heit 030 -009 0 R estrisik en @ 25\ mod _14 049 8331 557 7_6 8.do cx @ 1834 53 @ 2 @ 1
3.9 Remaining hazards
All CHRIST freeze-dryers were built state-of-the-art and according to the
accepted safety rules. Danger to life and limb of the operator, or of third
parties, or impairments of the units or other material assets, however,
cannot be completely excluded when the units are being used.
Use the freeze-dryer
• only for the purpose that it was originally intended for (see chapter 1.2 -
"Intended use") and
• only if it is in a perfect running state.
• Immediately eliminate any problems that can affect safety.
Pos: 78 / 010 Unive rsalm odul e/Absch nittswec hsel @ 0\m od_ 120 212 451 406 2_0. docx @ 41 8 @ @ 1