11.1 Mathematical relations

The automatic processes in the “Programmer module” menu (see chapter
6.5.3 - "Freeze-drying with the PGMplus programmer module") are based
on the following considerations:

Fig. 79: Graphical representation of the course of the control set value


Calculation of the control set value and of the gradient for the
temperature:

C/min]

[

sec

sec

sec

sec

tion

current

of

time

tion

tion

previous

of

value

set

tion

current

of

value

set

gradient

C]

[

sec

gradient

time

section

elapsed

tion

previous

of

value

set

value

set

control

Example:

Section

Set values

Section time [h:min]

Temperature [°C]

Preceding

...

30

Current

1:00

60

C/min

0.5

min

60

30

min

60

30

60

C

C

C

gradient

After an elapsed section time of 30 minutes, for example, the control set
value for the temperature is:

C

C/

.

C

value

set

Control

t

45

min

5

0

min

30

30

min)

30

(


Calculation of the control set value for the vacuum:

]

[

10

sec

.)

sec

.

(

10

.

sec

(

10

.)

sec

.

(

10

mbar

value

set

control

time

tion

elapsed

section

current

of

time

section

t

prev

value

set

LOG

t

current

value

set

LOG

t

prev

value

set

LOG

Pos: 235 /01 0 Univ ersal mod ule/Seit enwec hsel @ 0\m od_ 120 211 624 4312 _0. docx @ 10 5 @ @ 1

T/°C

t/h

1000

100

10

1

0.1

0.01

0.001

0.0001

+40

+30

+20

+10

0

-10

-20

-30

Pressure/mbar

Set value of the previous section

Course of the shelf temperature
control set value

Set value of the

current section

Previous

section

Current

section

Following

section