Martin Christ Delta 2-24 LSCplus User Manual
Page 93
Delta 1-24 LSC&
Delta 2-24 LSC&
10 Technical data
Version 04/2013, Rev. 1.2 of 20/02/2014 • sb
93
Translation of the original operating manual
Connection requirements
(without vacuum pump and
accessories)
Delta 1-24 LSC
plus
Delta 2-24 LSC
plus
Electrical connection:
3 x 400 V / 50 Hz
3 x 400 V / 50 Hz
Protection class:
I
I
IP protection category according to
DIN 60529:
11
11
Nominal power:
0.9 kVA
1.4 kVA
Nominal current:
3.5 A
3.5 A
Mains fuse:
10 A F
12 A F
Vacuum pump connection:
Depending on the type of vacuum
pump
Depending on the type of vacuum
pump
Filling quantity:
Delta 1-24 LSC
plus
Delta 2-24 LSC
plus
Refrigerant
See name plate
See name plate
Physical data (without vacuum
pump and accessories)
Delta 1-24 LSC
plus
Delta 2-24 LSC
plus
Dimensions
- Height:
- Width:
- Depth:
1,050 mm
860 mm
650 mm
(incl. vacuum flange connection)
1,050 mm
860 mm
650 mm
(incl. vacuum flange connection)
Weight:
approx. 215 kg
approx. 250 kg
Noise level according to DIN
45635:
__ dB(A)
__ dB(A)
EMC according to EN 55011:
Class B
Class B
Heat emission:
0.9 kW min.
1.9 kW max.
1.7 kW min.
3.0 kW max.
Equipment connections:
Delta 1-24 LSC
plus
Delta 2-24 LSC
plus
Vacuum connection:
Small flange connection DN25KF
(ISO 28403, DIN 2861)
Small flange connection DN25KF
(ISO 28403, DIN 2861)
Condensate drain valve:
Hose nozzle DN10
(outside diameter 12 mm)
Hose nozzle DN10
(outside diameter 12 mm)
Aeration valve:
Hose nozzle DN4
(outside diameter 8 mm max.)
Hose nozzle DN4
(outside diameter 10 mm max.)
Vacuum sensor:
SUB D-9
VSP 62
SUB D-9
VSP 62
Data interface (LAN):
RJ 45
RJ 45