Martin Christ Delta 2-24 LSCplus User Manual
Page 68
Delta 1-24 LSC&
Delta 2-24 LSC&
6 Operation
68
Version 04/2013, Rev. 1.2 of 20/02/2014 • sb
Translation of the original operating manual
• Pressing the button “Diagram” (3) displays the program in the form of a
diagram.
• Pressing the button “Program name” (2) calls up an input field in which
the name of the program can be changed.
• The button “Save” (7) can be used in between or at the end of the
program creation in order to save the program.
• The button “Program” (1) calls up the main window. The system will ask
the user whether the program shall be saved if this has not been done
yet.
The program will be automatically assigned to the first free program storage
location. The creation of the program is now complete and it can be loaded.
Button ">>more"
The button ">>more" can be used to enter more functions and set values in
a program-related manner.
Fig. 61: Selection of program-related functions and set values
Vacuum pump warm-up
See chapter 6.6 - "Optional extensions"
Option: Pressure increase test (only with the double chamber method)
(see chapter 6.6 - "Optional extensions")
The pressure increase test can only be performed when the freeze-dryer
is equipped with an intermediate valve.
NOTE
The indication of the measurement value "dp Test" can be configured in
chapter 6.5.1.1 - "Main window "Manual"" under "Actual values of the
current process" in the dialog box "Select actual value".