10 specifications – Laerdal Compact Suction Unit (LCSU) 4 User Manual
Page 15

10 Specifications
English
15
Power Source Options
Battery
Rechargable, NiMH 12 volt 1.6 Ah
DC Power-cord (12 V)
Dry Location Use Only
AC/DC Adapter Charger
Input: 100-240 V, 50-60 hz, 1.2 Ah
Output: +12 V, 3.4 Ah
Dry Location Use Only
External Battery Charger
Input: 110-240 V, 50-60 hz, 250 mA
Output: +18.5 V, 0.6 Ah
Dry Location Use Only
Environmental Conditions
Operating Temperature Range:
0 °C (+32 °F) – +40 °C (+104 °F)
Operating Relative Humidity:
0 - 95% (non condensing)
Operating Atmospheric Pressure:
10.2 Psi (70 kPA) – 15.4 Psi (106 kPA)
Storage & Transport Temperature:
-40 °C (-40 °F ) – +70 °C (+158 °F)
Storage & Transport Relative Humidity:
0 - 95% (non condensing)
Storage & Transport Atmospheric Pressure:
7.3 Psi (50 kPA) – 15.4 Psi (106 kPA)
Physical Characteristics
Dimensions
• 880051 (LCSU 4, 800 ml):
23.6 cm x 19 cm x 23.6 cm (9.3" x 7.5" x 9.3")
• 880061 (LCSU 4, 300 ml):
18.5 cm x 26.2 cm x 8.12 cm (7.3" x 10.3" x 3.2")
Weight
• 880051 (LCSU 4, 800 ml): 1.97 kg (4.35 lbs)
• 880061 (LCSU 4, 300 ml): 1.53 kg (3.375lbs)
Canister Capacity
• 300 ml
• 800 ml
Performance
Air flow @ Vacuum Inlet
All configurations: 30 LPM (free flow) typical (May be less
when running from internal battery)
Vacuum - Max.: 550+ mmHg
Vacuum - Range: 50 - 550+ mmHg
Vacuum Indicator accuracy: +/- 5% of full scale
Filter Specifications
Internal Filter of the 300 ml Disposable Canister
Porous plastic filter, PE material.
Internal Filter of the 800 ml Disposable Canister
Aerostate HEPA rated bacterial filter. White, unfinished
surface, cellulose with wet strength resin.
High Efficiency Filter Kit
With the High Efficiency Filter Kit installed the unit is in
accordance with ISO 10079-1. The Kit reduces the Air flow
and Battery run time. The filter is HEPA rated, hydrophobic,
with an efficiency of 99.97% down to a particle size of 0.3
μm.
Filter material: PTFE 1 μm
Water entry pressure: Min. 0.20 bar/150 mmHg/2.90 psi
(Flow to Canister)
Water entry pressure: Min. 0.87 bar/650 mmHg/12.57 psi
(Flow to Main Unit)
Airflow resistance: Max. 0.09 bar/67 mmHg/1.30 psi at 30
LPM Air flow
Nominal Filter diameter: 60 mm
Filter housing: Translucent or transparent
Material Chart
Cabinet: PC
Battery Cover:
PC
Vacuum Regulator:
PC
Bottom Cover:
PC
Vacuum Inlet Connector:
TPR
Control Panel:
PVC
800 ml Disposable Canister: GPPS
- Lid:
HDPE
- Internal Filter:
Aerostate
Vacuum Tube:
Silicone, K-Resin
Vacuum Port Connector:
TPR
Vacuum Inlet Connector:
PC
High Efficiency Filter:
PP
Filter housing:
K-Resin
300 ml Disposable Canister: PS
- Internal Filter:
PE
Patient Port:
PP
Patient Tube:
PVC
Wire Stands:
Steel, PVC
Electrical Requirements
100-240 VAC 50~60 Hz 1.2 A Max; 12 VDC, 3.4 Ah