7 Troubleshooting

English

12

Troubleshooting

Warning

Potential for Electrical shock. Do not attempt to open or disassemble pump or electrical accessories.

LCSU 4 will not power “ON”
1

Verify that the Battery is properly installed and charged.

2

Check alternative 12V DC or AC sources.

LCSU 4 runs, no Vacuum, or insufficient Vacuum
1

Verify proper Canister and Tubing connections.

2

Check system for leaks in Canister and Tubing connections.

3

Check Canister overflow protection float (800 ml).

4

Check if the Filter has been shut-off (300 ml).

Vacuum too high or too low.
Turn Vacuum Regulator to increase or reduce vacuum level.

Battery will not charge
1

Verify that the Battery is installed and connected.

2

Re-connect source power and observe charge (all connections).

Battery seems insufficient
1

Charge the Battery for 5 hours.

2

Perform the Battery Test.

Important
If the LCSU 4 condition is not resolved, contact Laerdal Medical or one of its authorized distributors for advice.