33

CHR xx

Change the instrument range according with the param-
eter value (xx):

• xx=00 pH range/0.001 resolution
• xx=01 pH range/0.01 resolution
• xx=02 pH range/0.1 resolution
• xx=03 mV range
• xx=04 Relative mV range

The instrument will answer for these commands with:

>

where:

is 02 ASCII code character (start of text)
is 03 ASCII code character (end of text)
:
is 06 ASCII code character (recognized command)
is 21 ASCII code character (unrecognized command)
is 24 ASCII code character (corrupted command)

COMMANDS REQUIRING AN ANSWER

The instrument will answer for these commands with:

where the checksum is the bytes sum of the answer string sent as 2 ASCII
characters.
All the answer messages are with ASCII characters.
RAS

Causes the instrument to send a complete set of readings in
according with the current range:

• pH, temperature and mV reading on pH range.
• Rel mV, absolute mV and temperature reading on

Relative mV range.

The answer string contains:

• Meter mode (2 chars):

• 00 - pH range (0.001 resolution)
• 01 - pH range (0.01 resolution)
• 02 - pH range (0.1 resolution)
• 03 - mV range
• 04 - Rel mV range

• Meter status (2 chars of status byte): represents a 8

bit hexadecimal encoding.
• 0x10 - temperature probe is connected
• 0x01 - new GLP data available
• 0x02 - new SETUP parameter