Hanna Instruments HI 2214 User Manual
Page 21

21
L O G G I N G F U N C T I O N
L O G G I N G F U N C T I O N
L O G G I N G F U N C T I O N
L O G G I N G F U N C T I O N
L O G G I N G F U N C T I O N
Up to 600 logged samples can be stored into memory.
A 100 manually logged records and 500 lot logging records can be
stored in the memory. To select logging type enter SETUP menu.
LOGGING THE CURRENT DATA
(manual logging)
Select the manual logging mode in SETUP menu.
To store the current reading into memory press
LOG while the instrument
is in measurement mode.
The instrument will display:
HI 2214 - the current date (mm.dd) on the primary LCD, the record
number on the secondary LCD and “
LOG” tag will blink for a few seconds
(see example below: record No. 15 dated September 14):
HI 2215 - “MAn” on the primary LCD, the record number on the secondary
LCD and “
LOG” tag for a few seconds (see example below: record No. 15):
followed by the number of free records:
If there are less than 6 memory locations remaining, the record number
and “
Lo” message will be displayed to alert the user.
If the log space is full (100 records), “
FULL LOG” message will be
displayed and no more data will be saved.
When
LOG is pressed, a complete set of information is stored: date, time,
pH, mV, temperature and calibration data.