Press the

ARROW keys to select another buffer to be

replaced.
Press

CFM to confirm the buffer that will be replaced.

Press

CAL to leave calibration without replacing.

Note: If the replaced buffer is outside the ±0.2 pH

window, around each of the calibrated buffers, it is
possible to select this buffer for next calibration
during current calibration.

WORKING WITH CUSTOM BUFFERS

If a custom buffer was set in SETUP menu, it can be selected during
calibration by pressing the

ARROW keys. The “BUFFER pH” tag will

blink.
Press

SETUP if you want to adjust the buffer value. The buffer value will

start blinking.
Use the

ARROW keys to change the buffer value.

After 5 seconds the buffer value is updated. Press

SETUP if you want to

change it again.

Note: Custom buffer value can be adjusted in a ±1.00 pH window,

around the set value.