SECOND CUSTOM BUFFER

Press

CAL when “cb2” is displayed. The custom buffer (disabled – “no”

or “

0” to “16” pH) and “CFM” will start blinking.

Press the

ARROW keys to change the custom buffer value.

Press

CFM to save the modified custom buffer value.

Press

CAL to escape without saving.

ONE-POINT CALIBRATION BEHAVIOR

Press

CAL when “1 Pnt” message is displayed on the secondary LCD.

One of the two options (“

Pnt” or “OFFS”) and “CFM” will start

blinking (see pH CALIBRATION PROCEDURE for details, page 9).

Press the

ARROW keys to toggle between “Pnt” and “OFFS” options.

Press

CFM to save the behavior for one-point calibration.

Press

CAL to escape without saving.

pH RESOLUTION

Press

CAL when “rES” message is displayed on the secondary LCD. The

set resolution (“

0.1“, “0.01“ or “0.001“) and “CFM” will start blinking.

Press the

ARROW keys to toggle between 0.1, 0.01 and 0.001 options

for

HI 2215; 0.1 and 0.01 options for HI 2214.

Press

CFM to save the modified value.

Press

CAL to escape without saving.

FIRST CUSTOM BUFFER

Press

CAL when “cb1” is displayed. The custom buffer (disabled – “no”

or “

0” to ”16” pH) and “CFM” will start blinking.

Press the

ARROW keys to change the custom buffer value.

Press

CFM to save the modified custom buffer value.

Press

CAL to escape without saving.

Note: To delete custom buffer enter change mode by pressing CAL then

press

CLR key.