• Use the

ARROW keys to set the reading on the secondary LCD to that

of the hot water.

When the reading is stable and close to the selected
calibration point, “

READY” tag will appear and

CFM” tag will blink.

• Press

CFM to confirm. The instrument memorize calibration and

restart to measurement mode.

Note: If the reading is not close to the selected calibration point,

WRONG” tag will blink. Change the temperature probe and

restart calibration.

m

m

m

m

mV CALIBRATION

V CALIBRATION

V CALIBRATION

V CALIBRATION

V CALIBRATION

(for technical personnel only)

All the instruments are factory calibrated for mV.
Hanna’s ORP electrodes are interchangeable and no mV calibration is
needed when they are replaced.
If the mV measurements are inaccurate, mV recalibration should be
performed.
For an accurate recalibration, contact your dealer or the nearest Hanna
Customer Service Center, or follow the instructions below.
A two-point calibration can be performed at 0.0 mV and 1800.0 mV.
• Attach to the BNC connector a mV simulator with an accuracy of

±0.1 mV.

• With the instrument off, press and hold down the

CAL & ºC

keys, then power on the instrument. The “

CAL” tag will appear and

the secondary LCD will show “

0.0 mV“.

• Set 0.0 mV on the simulator.

When the reading is stable and close to the selected calibration
point, “

READY” tag will appear and “CFM” tag will blink.

• Press

CFM to confirm. The secondary LCD will display

1800 mV

.

• Set 1800.0 mV on the simulator.

When the reading is stable and close to the selected calibration
point, “

READY” tag will appear and “CFM” tag will blink.

• Press

CFM to confirm. The

instrument

memorize calibration and

restart to measurement mode.