16

RELATIVE

RELATIVE

RELATIVE

RELATIVE

RELATIVE m

m

m

m

m V CALIBRATION

V CALIBRATION

V CALIBRATION

V CALIBRATION

V CALIBRATION

• Press

CAL when the instrument is in RELATIVE mV measurement

mode. The “

mV” and “ ” tags will be displayed. Absolute mV is

displayed on the primary LCD and “

AbS” message is displayed on the

secondary LCD.

• When the absolute reading is stable and in measurement range, the

instrument asks for confirmation.

• If the reading is out of range, “

WRONG” tag will be displayed.

• Press

CFM to confirm the absolute value. The instrument will display

0.0 mV“ on the primary LCD and “rEL” message on the secondary

LCD. In this moment the relative mV offset is equal to absolute mV
reading.

• Use the

ARROW keys if you want to change the displayed relative

mV value.

• Press

CFM to confirm the relative mV value. The relative mV

offset is displayed on the primary LCD. The instrument returns to
measurement mode.

Note: The relative mV value can be changed only inside the relative mV

offset window (± 2000 mV).