Setup mode allows viewing and modifying the following parameters:
• Expired Calibration Alarm
• Log Interval (
HI 2215 only)
• First Custom Buffer
• Second Custom Buffer
• One-point Calibration Behavior
• pH Resolution
• Current Time (hour & minute)
• Current Date (year, month & day)
• Beep Status
• Instrument Id
• Temperature Unit
To enter SETUP mode press
SETUP while the instrument is in
measurement mode.
Select a parameter with the
ARROW keys.
Press
CAL to change a parameter value. The selected parameter will
start blinking.
Press
RANGE to toggle between displayed parameters.
Press the
ARROW keys to increase or decrease the displayed value.
Press
CFM to save the modified value or CAL to escape without saving.
LOG INTERVAL
(HI 2215 only)
Press
CAL when log interval is displayed. The log interval and “CFM” will
start blinking (“
MAn” for log on demand, “StAb” log on stability,
interval in seconds or minutes).
Press the
ARROW keys to change the custom buffer value.
Press
CFM to confirm the selection.
Press
CAL to escape without saving.
EXPIRED CALIBRATION ALARM
Press
CAL when the calibration time-out is displayed. Calibration time-out
(“
OFF” or “1” to “7” days) and “CFM” will start blinking.
Press the
ARROW keys to change the calibration time-out value.
Press
CFM to save the modified calibration time-out value.
Press
CAL to escape without saving.