Hanna Instruments HI 2214 User Manual

Page 30

background image

30

All the instruments are factory calibrated for temperature.
Hanna’s temperature probes are interchangeable and no temperature
calibration is needed when they are replaced.
If the temperature measurements are inaccurate, temperature recalibration
should be performed.
For an accurate recalibration, contact your dealer or the nearest Hanna
Customer Service Center, or follow the instructions below.
• Prepare a vessel containing ice and water and another one containing

hot water (around 50 ºC). Place insulation material around the
vessels to minimize temperature changes.

• Use a calibrated thermometer with a resolution of 0.1ºC as a reference

thermometer.

• With the instrument off, press and hold down the

CFM & SETUP

keys, then power on the instrument. The “

CAL” tag will appear and

the secondary LCD will show “

0.0 ºC“.

• Submerse the temperature probe in the vessel with ice and water as

near as possible to the reference thermometer. Allow a few seconds for
the probe to stabilize.

• Use the

ARROW keys to set the reading on the secondary LCD to that

of ice and water, measured by the reference thermometer. When the
reading is stable and close to the selected calibration point, “

READY”

tag will appear and “

CFM” tag will blink.

• Press

CFM to confirm. The secondary LCD will show

50.0

º

C

.

• Submerse the temperature probe in the second vessel as near as

possible to the

reference thermometer. Allow a few seconds for the

probe to stabilize.

TEMPERATURE CALIBRATION

TEMPERATURE CALIBRATION

TEMPERATURE CALIBRATION

TEMPERATURE CALIBRATION

TEMPERATURE CALIBRATION

(for technical personnel only)