Notes: • If the value measured by the instrument is not close to the

selected buffer, “

WRONG” “ ” and “WRONG” “ ” tags

will blink alternately. Check if the correct buffer has been
used, or clean the electrode by following the Cleaning
Procedure (see page 38). If necessary, change the buffer or
the electrode.

• When a custom buffer is displayed, the “

BUFFER pH” tag

blinks. To change the custom buffer value in accordance
with the buffer temperature proceed as described in
“WORKING WITH CUSTOM BUFFERS” (see page 14).

• If the buffer temperature or the manual temperature

exceeds the temperature limits of the buffer, “

WRONG” tag

and temperature reading will blink.

• If “

WRONG”, “BUFFER pH” tags and “OLd” message are

displayed blinking on the secondary LCD line, an inconsistency
between new and previous (old) calibration is detected.
Clear calibration parameters and proceed with calibration
from the current calibration point. The instrument will keep
all confirmed values during current calibration.

• To clear calibration parameters for all buffers starting with

current buffer, press

CLR. The calibration will continue from

the current point. If this procedure is performed while
calibrating in the first point, the instrument returns to
measurement mode.

• Press

RANGE to toggle between pH

buffer, calibration buffer number and
temperature reading.

• Each time a buffer is confirmed, the new calibration data

replaces the old calibration data for the corresponding buffer.
If current buffer has no previous data in the existing
calibration and the calibration is not full (five buffers), the
current buffer is added to the existing data. If the existing
calibration is full, the instrument asks which buffer to be
replace.